Clean Air EquipmentRegenerative Medicine Cabinet

This equipment ensures the cleanliness of the working chamber and also guarantees operator safety and sample protection when cells are handled.
* The product cannot be used for the handling of pathogenic bacteria.

Features (Models RCV-1900 and RCV-1300)

Achieves both grade A (ISO Class 5) cleanliness and airflow balance performance

  • Laminar airflow within the working chamber (outlet wind speed: 0.45 m/s ± 20%)
  • Performance guaranteed using an airflow balance test: (1) Operator safety test, (2) Sample protection test

Simplified clean booth/incubator (cell culture apparatus) and others can be connected.

A monitor can be installed in front of an operator.

A garbage chute (waste container) and a cell observation device can be installed (garbage chutes cannot be installed in RCV-1300.)

  • The RCV-1300C is also available for when the entrance for bringing equipment into the room is too small.

Table of specifications

Model RCV-1900 RCV-1300
Dust collection parts HEPA filter
Dust collection efficiency Captures 99.99% or more for 0.3-µm particles. The product has passed a scan test.
Front shutter opening height 200mm
Cleanliness in the working chamber ISO Class 5 (Class 100 in the old US federal standard, when the front shutter is opened 200 mm)
Air volume Exhaust 14.0-16.4 m3/min
(Center value: 15.2 m3/min)
9.4 - 10.9m3/min
(Center value: 10.1 m3/min)
Air velocity Downflow Average: 0.45 m/s ± 20%
Immediately below the outlet surface: 150 mm, measured values at different points are within ± 20% of the average outlet wind speed
Average: 0.34-0.47m/s Average: 0.32-0.42m/s
(100 mm above the lower end of the front shutter)
Inflow Average: 0.60-0.70 m/s
Circulation frequency 1920-2880 times/h (central value: 2400)
Airflow balance test Methods Spores of Bacillus subtilis are sprayed using a nebulizer and samples are taken.
Operator safety test Number of colonies collected from synthesized suspension based on air collected with an AGI sampler: 10 or less
Number of colonies collected from a slit sampler: 5 or less
Sample protection test Number of colonies in Petri dishes placed on the working table: 5 or less
Inter-sample cross-contamination prevention test N/A
Airtightness of the main body The pressure decreases 10% or less 30 minutes after the inside of the main body is pressurized at 500 Pa (the positive pressure maintenance method)
Power supply AC, 1φ, 100 V, 50Hz or 60Hz
Power consumption 690W±15% 390W±15%
(compatible with both 50 and 60 Hz), (excluding connections with optional equipment)
Rated working current value (A) 8.4A±15% 5.0A±15%
(compatible with both 50 and 60 Hz), (excluding connections with optional equipment)
Outer dimensions W2,370×D1,170×H2,340 W1,620×D1,170×H2,340
Height at delivery (mm) 2,280
Main body weight Approx. 510kg Approx. 390kg

Features (Model RCV-1300C, compact type)

Achieves both grade A (ISO Class 5) cleanliness and airflow balance performance

  • Performance guaranteed using an airflow balance test: (1) Operator safety test, (2) Sample protection test

Simplified clean booth/incubator (cell culture apparatus) and others can be connected.

A monitor can be installed in front of an operator.

Can be carried through 800 mm-wide doors

Specification table (Model RCV-1300C, compact type)

Model RCV-1300C
Dust collection parts HEPA filter
Dust collection efficiency Captures 99.99% or more for 0.3-µm particles. The product has passed a scan test.
Cleanliness in the working chamber ISO Class 5 (Class 100 in the old US federal standard, when the front shutter is opened 200 mm) Class of surroundings: ISO Class 8 (or Class 100,000)
Front shutter opening height 200mm
Air volume Exhaust 8.6 - 10.1 m3/min
Air velocity Downflow Average: 0.30-0.40 m/s
Inflow Average: 0.55-0.65 m/s
Airflow balance test Methods Spores of Bacillus subtilis are sprayed using a nebulizer and samples are taken.
Operator safety test Number of colonies collected from synthesized suspension based on air collected with an AGI sampler: 10 or less
Number of colonies collected from a slit sampler: 5 or less
Sample protection test Number of colonies in Petri dishes placed on the working table: 5 or less
Inter-sample cross-contamination prevention test N/A
Airtightness of the main body The pressure decreases 10% or less 30 minutes after the inside of the main body is pressurized at 500 Pa (the positive pressure maintenance method)
Power supply AC, 1φ, 100 V, 50Hz or 60Hz
Power consumption 280 W ± 15% (compatible with both 50 Hz and 60 Hz), (excluding connections with optional equipment)
Maximum 310 W during accelerated operation when HEPA filters are clogged
Rated working current value 3.7 A ± 15% (compatible with both 50 Hz and 60 Hz), (excluding connections with optional equipment)
Maximum 4.1 A during accelerated operation when HEPA filters are clogged
Outer dimensions W1,620×D900×H2,180mm
Dimensions inside the working chamber W1,300×D600×H675mm
Maximum dimensions at delivery W1,620×D795×H1,970mm
(with cover for monitor installation on the rear side detached)
Main body weight Approx. 300kg