Clean Air EquipmentBiological Airlock Room

This equipment is installed between a contaminated zone and a non-contaminated zone (a general room or corridor). It uses negative pressure to suppress the dispersion of contaminants.

Features

Airtight doors for entry/exit

Doors with an interlock device

Interior pressure can be checked using a differential pressure gauge

HEPA filters for air supplying air to and exhausting air from the airlock are replaced from inside the room

Table of specifications

Model BHR-800A
Dust collection parts Pre-filter: Nylon non-woven fabric; Main filter: HEPA filter (H-110A)
Dust collection efficiency (%) Captures 99.99% or more for 0.3-µm-diameter particles. The product has passed a scan test.
Rated air volume (m3/min) 1 (Ventilation frequency: Approx. 20 times/hour)
Main Unit Made of steel sheet (based on Bonderized steel sheet ( T1.2 - T1.6)
Door Made of steel sheet with a wired glass window (T6.8) and includes a special handle
Floor Hairline finished SUS304 sheet
Coating color Neo White semi-gloss (equivalent to Munsell No. 5Y 8.5/1)
Power supply Single-phase AC 100 V, 50Hz or 60Hz
Illuminating lamp 1 20 W lamp (daylight color)
Germicidal lamp 1 15 W lamp
Internal pressure Please set it to the midpoint between the contaminated and non-contaminated areas.
(A differential pressure gauge will display the internal pressure.)
Door interlock mechanism When one door is opened, an electromagnetic lock prevents the other door from being opened.
Delivered form This equipment can be divided into two parts (It can be divided for delivery, but this requires on-site sealing.)
Product mass (kg) 650

"Clean Air Equipment" Bio Safety Cabinets

"Clean air equipment" Select from lineup

Top page of "Clean air equipment"

Catalog Lists and Inquiries

Click the links below to make inquiries about products, request a quotation, and download catalogs. Please feel free to consult with us.