- Home
- Products and Solutions
- Clean Air Equipment
- Biological Airlock Room
Clean Air EquipmentBiological Airlock Room
This equipment is installed between a contaminated zone and a non-contaminated zone (a general room or corridor). It uses negative pressure to suppress the dispersion of contaminants.

Features
Airtight doors for entry/exit
Doors with an interlock device
Interior pressure can be checked using a differential pressure gauge
HEPA filters for air supplying air to and exhausting air from the airlock are replaced from inside the room

Table of specifications
| Model | BHR-800A |
|---|---|
| Dust collection parts | Pre-filter: Nylon non-woven fabric; Main filter: HEPA filter (H-110A) |
| Dust collection efficiency (%) | Captures 99.99% or more for 0.3-µm-diameter particles. The product has passed a scan test. |
| Rated air volume (m3/min) | 1 (Ventilation frequency: Approx. 20 times/hour) |
| Main Unit | Made of steel sheet (based on Bonderized steel sheet ( T1.2 - T1.6) |
| Door | Made of steel sheet with a wired glass window (T6.8) and includes a special handle |
| Floor | Hairline finished SUS304 sheet |
| Coating color | Neo White semi-gloss (equivalent to Munsell No. 5Y 8.5/1) |
| Power supply | Single-phase AC 100 V, 50Hz or 60Hz |
| Illuminating lamp | 1 20 W lamp (daylight color) |
| Germicidal lamp | 1 15 W lamp |
| Internal pressure |
Please set it to the midpoint between the contaminated and non-contaminated areas.
(A differential pressure gauge will display the internal pressure.) |
| Door interlock mechanism | When one door is opened, an electromagnetic lock prevents the other door from being opened. |
| Delivered form | This equipment can be divided into two parts (It can be divided for delivery, but this requires on-site sealing.) |
| Product mass (kg) | 650 |
"Clean Air Equipment" Bio Safety Cabinets
Catalog Lists and Inquiries
Click the links below to make inquiries about products, request a quotation, and download catalogs. Please feel free to consult with us.